DATAQ DI-808 Web-based Voltage and Thermocouple Data Logger system. Features 8 isolated analog input channels programmable as a thermocouple or voltage input, 16-bit measurement resolution, and a full scale range of ±10mV to ±50V for voltage measurements or Thermocouple types J, K, T, B, R, S, E, or N.
Its built-in web server allows you to access, configure, manage, and oversee in real time all aspects of the instrument’s measurement process using any device with a web browser running under any operating system. Just as you can navigate to Yahoo! using your favorite smart device, that same device can be used without exception with the DI-808. And because intelligence is built into the instrument, third party cloud services with their associated security risks and fee-based structures are avoided entirely. You stay in control at all levels, from device configuration and real time monitoring, to the acquired data and alarms the DI-808 places on your smart device. The DI-808’s Ethernet interface allows it to integrate with any existing local area network (LAN) or virtual private network (VPN.) And simple port forwarding exposes the instrument to the Internet, allowing remote access from any location on the planet.
The DI-808 also provides four alarm output channels that can be used to flag exceptions during a recording. Each discrete output can be used to signal a PLC, turn on an alarm, or otherwise used as demanded by the application.
The DI-808 supports exceptional triggering flexibility to determine when the data logger should start and stop recording data, and when alarms should be activated.
- Includes a DI-808 instrument, a screwdriver for signal connections, a power supply
- No software to install
- Stores data locally to internal flash memory or a USB thumb drive and/or remotely to any accessible server or email address
- Allows remote real time display and configuration
- Eight differential and isolated analog input channels
- 16-bit analog measurement resolution
- Four pulse input channels programmable as count, frequency, or discrete function per channel
Specifications
Analog Inputs
- Number of Channels: 8
- Configuration: Differential, Isolated
- Isolation: Input-to-output, channel-to-channel
- Programmable voltage measurement ranges per channel: ±10 mV, ±25 mV, ±50 mV, ±100 mV, ±250 mV, ±500 mV, ±1 V, ±2.5 V, ±5 V, ±10 V, ±25 V, ±50 V
- Voltage measurement accuracy: ±(0.05% of range + 10 µV)
- Input impedance: 1 MΩ, all ranges
- Absolute maximum input without damage: 120 V (±dc or rms)
- Maximum common mode voltage: 120 V (±dc or rms)
- Min common mode rejection (330Ω unbalance): >110 dB (DC to 60 Hz)
- Alarm and trigger hysteresis:
- Voltage: ±0.5% of the full scale range
- Temperature: ±0.3 °C
Digital/Pulse Inputs
- Number of Channels: 4
- Pull-up value: 4.7 KΩ
- Isolation: None
- Input high voltage threshold: 1.8 V minimum
- Input low voltage threshold: 1.4 V maximum
- Absolute voltage input without damage: 0 ≤ V ≤ 30 V
- Maximum count value:
- 232 – 1 (.csv format)
- 65,535 (WinDaq format)
- Maximum measured frequency:
- >100 kHz (.csv format)
- 65,565 Hz (WinDaq format)
Alarm Outputs
- Number of channels: 4
- Maximum drain voltage: 30 V
- Maximum sink current: 100 mA
Internal Memory
- Type: Non-volatile flash
- Size: 4 GB (standard); 32 GB (optional)
- Use: Automatic backup of recorded data
Power I/O
- Input power requirements:
- 9-36 V dc @ 7.5 Watts
- 9-36 V dc @ 10 Watts (w/excitation)
- Excitation power supply output: 15 V, 130 mA max, 13 mA min
Configurable Components
- Network variables: IP address, Subnet mask, Gateway, DNS
- Account log-on information: For FTP, SFTP, SMB, SMTP
- User log-on information: User and administrator names and passwords
Supported Standards
- Network Time Protocol (NTP): Syncs internal time and date clock to Internet time
- File Transfer Protocol (FTP): Allows the instrument to push re-corded data to an FTP server
- Secured File Transfer Protocol (SFTP): Allows the instrument to push re-corded data to an SFTP server
- Server Message Block (SMB) protocol: Allows the instrument to record data to a local server drive
- Simple Mail Transfer Protocol (SMTP): Allows the instrument to send data and alarms to multiple email addresses
Related Products
DI-4718B-U Ethernet Data Logger and Data Acquisition System for applications that require signal conditioning. Install up to eight DI-8B isolated signal conditioning modules on the...
Ethernet data acquisition system with stand-alone capability (records to SD card) featuring 8 inputs for DI-8B signal conditioning modules (not included), a maximum throughput rate...
USB data acquisition system with stand-alone capability (records to SD card) featuring 8 inputs for DI-8B signal conditioning modules (not included), a maximum throughput rate...
DATAQ Instruments announces model DI-788, a new 32-channel data acquisition system designed specifically to accept DI-8B style modular signal conditioners and based on our Third...