Feautures:
- Self-contained calibration system
- Test amplitude of 1 m/s² rms
- Vibration frequency from 70 Hz to 10 kHz
- Suitable for test objects weighing up to 400 gr.
- Internal signal conditioner for charge, IEPE and voltage
- Traceable to PTB standard
- USB interface for PC controlled calibration
- PC software for calibration and display with sensor data base included
- Customer software control by simple ASCII commands
- Mains buffered NiMH battery for desktop and field use
- Unequalled price-performance ratio
Specifications
| Shaker System | |
| Vibration acceleration (rms) | 1 m/s² |
| Vibration frequency | 70 .. 10 000 Hz |
| Weight of test object max. | 400 g |
| Basic accuracy at 23°C ± 10 °C | at 80 / 160 / 315 / 630 Hz: ± 2 % calibrated
at 70 .. 1000 Hz: ± 2 % typically at 1000 .. 10 000 Hz ± 5 % typically |
| Harmonic Distortion (THD) | < 2 % (70 .. 3000 Hz) |
| Transverse vibration | < 10 % (70 .. 400 Hz; m < 200 g) |
| Temperature coefficient max. | ±0,02 %/K |
| Sensor mounting | M5 thread, magnet |
| Signal Conditioner | |
| Charge input | Single-ended, Umax= 500 V, BNC socket at rear |
| Voltage input / IEPE | Single-ended, 100 MΩ, Umax= 500 V, BNC socket at rear, alternatively with IEPE constant current supply, ICONST = 3.8 .. 5.6 mA, US > 22 V |
| Signal conditioner ranges | Autoranging, 0 / 20 / 40 / 60 dB or 1 / 10 / 100 / 1000 mV/pC |
| Transducer sensitivity range | 0.12 mV/ms-2 .. 1.2 V/ms-2 (voltage); 0.12 pC/ms-2 .. 1.2 nC/ms-2 (charge) |
| Readout display | Transducer sensitivity, selectable between metric or English engineering units |
| Accuracy | Voltage / IEPE: ± 0.3 %; Charge: 0.6 %; 16 Bit ADC resolution |
| General | |
| Temperature range | -10 .. 55 °C |
| Operating time per battery charge | 2 .. 4 h depending on duration of shaker use |
| Accumulator charging time | 2 h |
| PC serial interface | USB 2.0 with Minii-B socket |
| Dimensions | 205 x 105 x 90 mm |
| Weight | 3 kg |
| Standard accessories | Transport case, mains adapter (100 .. 240 VAC, 50 / 60 Hz, PC software, USB cable, thread adapters (M5, M8, 1/4″-28, UNF 10-32), Adapter BNC / UNF 10-32 (Mod. 017), silicon grease |
Q&A
Related Products
Low-cost accelerometer IEPE output Includes electronic data sheet (TEDS) For light test objects Insulated case M5 base thread UNF 10-32 socket (Microdot)
Vibration accelerometer KSI 80VC/82VB with 4-20mA output Easy to connect to PLC or indicators with 4-20mA output Intergrated signal conditioning function for harsh enviroments or...
Rapid and easy calibration of vibration transducers for acceleration, velocity and displacement Calibration and fault detection of vibration measuring system
General purpose shear-type accelerometer Two sensitivity versions Low sensitivity to temperature transients Low influence of base bending effects IEPE compatible output allows long cables High...