- Load independent vibration level for test objects up to 600 grams
- Vibration magnitude of 10 m/s² (VC20) or selectable in up to 5 steps from 1 to 20 m/s²
- Quartz controlled vibration frequency of 159.15 Hz (VC20) or selectable in seven steps from 15.92 Hz to 1280 Hz (VC21/VC21D)
- Display shows frequency, magnitude, error in percent and calibration date
- VC21D with detachable fastening device for the dynamic calibration of non-contact displacement sensors / proximity probes
- VC21D with clock output for the synchronization of narrow-band measuring systems
- Traceable to national PTB standard, calibration to ISO 16062-44, factory calibration certificate included
- Rugged design
- Mains buffered battery operation for laboratory and field use
- Mains plug adapter with wide range input included
Specifications
| VC21 / VC21D | |||||||||||||
| Vibration frequencies | 15,92 Hz | 40 Hz | 80 Hz | 159,2 Hz | 320 Hz | 640 Hz | 1280 Hz | ||||||
| Vibration accelerations (rms) | 1 m/s² | 1 m/s² | 1 m/s² | 1 m/s² | 1 m/s² | 1 m/s² | 1 m/s² | ||||||
| 2 m/s² | 2 m/s² | 2 m/s² | 2 m/s² | 2 m/s² | 2 m/s² | 2 m/s² | |||||||
| 5 m/s² | 5 m/s² | 5 m/s² | 5 m/s² | 5 m/s² | 5 m/s² | ||||||||
| 10 m/s² | 10 m/s² | 10 m/s² | 10 m/s² | 10 m/s² | |||||||||
| 20 m/s² | 20 m/s² | 20 m/s² | 20 m/s² | ||||||||||
| Vibration velocities (rms) at radian frequencies 100 Hz and 1000 Hz | 10 mm/s | 1 mm/s | |||||||||||
| 20 mm/s | 2 mm/s | ||||||||||||
| 5 mm/s | |||||||||||||
| 10 mm/s | |||||||||||||
| 20 mm/s | |||||||||||||
| Vibration displacements (rms) at radian frequencies 100 Hz and 1000 Hz | 100 µm | 1 µm | |||||||||||
| 200 µm | 2 µm | ||||||||||||
| 5 µm | |||||||||||||
| 10 µm | |||||||||||||
| 20 µm | |||||||||||||
| Weight of test object, max.
1 m/s² |
500g |
500g |
500g |
500g |
500g |
500g |
500g |
||||||
| 2 m/s² | 500g | 500g | 500g | 500g | 500g | 500g | 500g | ||||||
| 5 m/s² | 500g | 500g | 500g | 500g | 500g | 500g | |||||||
| 10 m/s² | 500g | 500g | 500g | 400g | 200g | ||||||||
| 20 m/s² | 250g | 200g | 100g | 50g | |||||||||
| Weight of test object, max. | 500 g (1 .. 10 m/s²); 200 g (20 m/s²) | ||||||||||||
| Amplitude error, max | ± 3 % (0 .. 40 °C / 32 .. 100 °F)
± 5 % (-10 .. 55 °C / 14 .. 130 °F) |
||||||||||||
| Frequency error, max. | ± 0,05 % | ||||||||||||
| Clock output (VC21D) | Clock of internal reference sensor; BNC; 3.3 V; 50 Ω; duty ratio approx. 1:1 | ||||||||||||
| Transverse vibration (14 mm above shaker) | < 10 % | < 10 % | < 10 % | < 10 % | < 20 % | < 20 % | < 10 % | ||||||
| Harmonic distortion | < 1 % | < 5 % | < 1 % | < 1 % | < 1 % | < 1 % | < 1 % | ||||||
| Level indication | percent display and piezo beeper | ||||||||||||
| Sensor mounting | M5 tapped hole (90° ± 1°; 7mm deep), magnet | ||||||||||||
| Operating temperature range | -10 .. 55 °C | ||||||||||||
| Accumulator type | built-in NiMH battery pack; 7.2 V / 1.6 Ah | ||||||||||||
| Operating time per battery charge | approx. 5 hours with 100 g weight | ||||||||||||
| Charge socket | Circular power connector to DIN 45323 (5.5 mm / 2.2 mm) | ||||||||||||
| Charging time | approx. 4 hours | ||||||||||||
| Charge condition indication | bar graph display | ||||||||||||
| Charging voltage | 11 .. 18 VDC | ||||||||||||
| Charging current | < 1 A | ||||||||||||
| Automatic switch off | 1 .. 30 Minutes adjustable | ||||||||||||
| Dimensions (L x W x H) | 100 x 100 x 120mm | ||||||||||||
| Weight | 2,2 kg | ||||||||||||
| Accessories (scope of delivery) | Plastic carrying case, mains adapter PS1600 (100 .. 240 VAC; 50 / 60 Hz), thread adapters (M3, M5, M8, 1/4″-28, UNF 10-32)
|
||||||||||||
| Thread adapters for VC21D (optional) | M6x0,5; M8x1; M10x1; M14x1; M20x1; 1/4″-28; 3/8″-24; 1/2″-20 | ||||||||||||
Q&A
Related Products
Low-cost accelerometer IEPE output Includes electronic data sheet (TEDS) For light test objects Insulated case M5 base thread UNF 10-32 socket (Microdot)
Versatile tool for vibration measurement during product development and for health and safety at the workplace Measurement of hand-transmitted vibration to ISO 5349 and ISO/TR...
Vibration accelerometer KSI 80VC/82VB with 4-20mA output Easy to connect to PLC or indicators with 4-20mA output Intergrated signal conditioning function for harsh enviroments or...
Self-contained calibration system Test amplitude of 1 m/s² rms Vibration frequency from 70 Hz to 10 kHz Suitable for test objects weighing up to 400...