T-Guard2 Leak Detection Sensor – INFICON

Brand: INFICON
Product code: T-Guard2

    INFICON T-Guard2 Leak Detection Sensor closes the gap between costly hard vacuum helium leak detection and low sensitivity leak testing methods like waterbath and pressure decay.

    T-Guard2 works at atmospheric pressure, therefore there is no need for vacuum chambers and pumps.

    There is no faster and more cost-effective way of repeatable leak testing in the measurement range of T-Guard2.

    ADVANTAGES

    • Low operating cost
      • The INFICON Wise Technology Helium sensor does not need a vacuum, turbo-pump, or maintenance. This lowers your operating costs and assure high uptimes.
    • No limitation of product properties
      • Even large, warm or humid objects, and those can not be well operated under vacuum can be leak tested
    • Find all leaks reliably
      • With INFICON Wise Technology leaks down to 1×10-6 mbar·l/s can be detected reliably
    • Easy to use
      • The intelligent Software and optional display units supply easy operation through an intuitive menu structure
    • Fexible Control
      • INFICON T-Guard2 allows a wide range of control options. You can choose between PLC, PC or optional display and Profibus.

    TYPICAL APPLICATIONS

    • Leak detection for air conditioner components
    • Leak detection for components for the automotive industry
    • Leak detection for large, warm or humid components
    • Leak detection for all components where pressure decay and water bath systems can not provide sufficient sensitivity
    • Other markets where helium vacuum leak detection has been considered as too costly or complex
    T-Guard2 Leak Detection Sensor
    Part Number Description
    540-200 WISE Technology T-Guard2 Sensor
    540-201 WISE Technology T-Guard2 Sensor Profibus

    ACCESSORIES

    T-Guard2 Leak Detection Sensor
    Part Number Description
    551-100 Control unit, desktop version
    551-101 Display unit, rack installation
    551-102 Cable 5 m for 551–100
    551-103 Connecting cable, ext. Control unit, 0.7
    590-571 Sensistor ILS500 F Leak Detection Filler
    590-573 Sensistor ILS500 F-HP Leak Detection Filler, high pressure model

    Specifications

    SPECIFICATIONS

    Type T-Guard2 Leak Detection Sensor
    Minimum detectable leak rate mbar•L/s 1E-6
    Measurement range decades 5
    Test chamber pressure Atm 1
    Operation modes 2 automatic, 1 advanced
    Maximum carrier gas flow sccm 1.000.000
    Probe gas flow FINE/GROSS sccm 180 / 90
    Time constant of the leak rate signal s <1
    Run up time min <3
    Control inputs SPS compatible (max.35V)
    Status / Trigger outputs 8 x relaiscontacts (max. 60V DC / 25V AC / 1A)
    recorder output lin/log 0 – 10V
    Recommended fore pump Membranepump, two-stage
    Power supply demand V (dc) 24
    Typical power consumption Watt 100
    Hose connectors mm 6
    Type of protection IP20
    Dimensions mm 272 x 130 x 272
    Weight kg 6,8

    Datasheet

    Related Products

    Modul1000 Helium Leak Detector - INFICON

    The Modul1000 is a plug-and-play helium leak detector for universal integration into industrial leak testing systems (vacuum or sniffer mode)

    Product detail
    UL3000 Fab ULTRA NEW - INFICON

    Specifically designed for the needs of leak checking in semiconductor applications. It features flexibility, mobility, fast start up, high sensitivity, fast and precise test results,...

    Product detail
    XL3000flex Helium and Hydrogen Sniffer Leak Detector - INFICON

    HIGH FLOW Technology for Highest Measurement Certainty and Fastest Throughput