The INFICON passive Cold Cathode Gauge Heads MAG050 and MAG060 are designed for use with the INFICON Vacuum Gauge Controller VGC083C.
They are constructed of a compact metal design resulting in a simple yet rugged gauge suitable for numerous industrial applications. Gauge Head MAG050 utilizes an elastomer internal seal while the MAG060 gauge head uses a metal seal allowing pressure measurements in the UHV range.
The Gauge Head assembly can be easily disassembled and cleaned allowing Long term use with minimal down time. A Gauge Head ignition aid mounted on the anode improves the time it takes to set the Cold Cathode Gauge Head on.
MAG050 measures down to 2×10-9 mbar while MAG060 is able to measure down to 1×10-10 mbar.
MAG050 is bakeable up to 150°C and it is mainly intended for use in different industrial applications.MAG060 bakeable up to 250 °C and radiation resistant, which makes it an excellent choice for for the use in R&D applications.
Specifications
| MAG050 | MAG060 | |
| Measurement system | cold cathode ionization measurement system (according to the inverted magnetron principle) | |
| Measurement range (air, N2) hPa | mbar | 2×10-9 … 5×10-3 | 1×10-10 … 5×10-3 |
| Accuracy (N2, typical) % of reading | 30 | |
| Repeatability (typical) % of reading | 5 | |
| Overpressure kPa (bar) | £900 | |
| Operation temperature (°C) | +5 … +80 (with normal cable)
+5 … +150 (with high temperature cable) |
+5 … +80 (with normal cable)
+5 … +250 (with high temperature cable) |
| Storage temperature ( °C) | –40 … +80 | –40 … +80 |
Q&A
Related Products
The NEW INFICON Vacuum Gauge Controller VGC083 is designed for use with passive gauge heads BAG05x and PGE050 in a fixed combination of two PGE050...
Measuring range: 10… 1000 Torr (mbar) Accuracy: 1% of scale (0.5% optional) Resolution: 0.05% of scale Minimum reading: 0.05% of scale Temperature compensation range: 10…...
True high precision pressure measurement – Ceramic technology Full stable output – Proven by PTB Flexible communication – Various modern interfaces All functions integrated –...
The INFICON Bayard-Alpert passive vacuum gauge heads BAG050, BAG051, BAG052 and BAG053 are designed for use with the INFICON Vacuum Gauge Controller VGC083A & VGC083B....