Industrial USB/Ethernet Data Acquisition and Data Logger DI-4730

Brand: DATAQ
Product code: DI-4730

    The DI-4730 is the next-generation of our very popular DI-730 series. The two are functionally identical except that the DI-4730 supports both ChannelStretch™ channel expansion and stand-alone data logging to a removable USB flash drive (unlock code required). Suitable for a wide range of analog and discrete measurements, the DI-4730, in tandem with the included WinDaq software, allows real time data acquisition, playback, and analysis performance that covers an incredibly broad array of applications in both laboratory and industrial settings. With all eleven scan list items enabled (8 analog channels, one each counter and rate, and all discrete input ports in the remaining position) a total sample throughput rate of 220 kHz is supported.

    • Measurement range of ±10mV to ±1000VDC (or peak AC) over six ranges – ±1000V, ±100V, ±10V, ±1V, ±100mV, ±10mV full scale
    • Stand-alone capability to USB thumb drive (WinDaq/Pro Unlock Code required to record all channels to USB thumb drive – See Accessories)*
    • Up to 160,000 Hz Sample Throughput Rate using WinDaq Acquisition software
    • Up to 160,000 Hz Sample Throughput Rate in stand-alone mode using high speed microSD card
    • 12 to 16-bit analog-to-digital resolution
    • Power over Ethernet (PoE) included – run the device via USB, Ethernet, or PoE
    • Adaptable low-pass filter per channel
      • Corner frequency set as a function of sampling rate
    • Seven digital inputs
      • Support for rate measurements to 50 kHz
      • Support for 16-bit counter
    • Supports ChannelStretch™ multi-unit synchronization

    Specifications

    Analog Inputs

    • Number of Channels: 8
    • Configuration: Differential
    • Full Scale Range: Programmable ±10mV, ±100mV, ±1V, ±10V, ±100V, ±1000V
    • Isolation: ±1000V input-to-output and channel-to-channel
    • Max normal mode voltage (VNM): 800VDC or peak AC
    • Max common mode voltage (VCM): 1000VDC or peak AC
    • Max Input without Damage: ±1000VDC or Peak AC
    • Common Mode Rejection: 100dB min @ DC to 60Hz
    • Accuracy:
      • Vin ≤ 800V ±(0.25% of full scale range ±100µV)
      • Vin > 800V ±(0.5% of full scale range)
    • Input impedance: 10MΩ for all ranges

    ADC Characteristics

    • Resolution: 12- to 16-bit
    • Max. throughput sample rate:
      • Standard: 160 kHz
      • Stand-alone: 20 kHz to 160 kHz (dependent on flash memory)
    • Min. throughput sample rate:
      • Standard: 20 Hz (2.2 samples per hour with WinDaq)
      • Stand-alone: 0.0017 Hz
    • Sample rate timing accuracy: 100ppm

    Digital Inputs

    • Number of ports: 7
    • Configuration: Inputs
    • Threshold levels: TTL-compatible
    • Absolute maximum applied voltage (V): 25 V dc
    • Reserved Digital Inputs
      • D0: WinDaq Remove Events
      • D1: WinDaq Remote start/stop
      • D2: Rate Input
      • D3: Count Input
      • Count/Rate Inputs
    • Internal pull-up value: 4.7 kΩ
    • Input high voltage threshold: 2.4 V
    • Input low voltage threshold: 0.8 V
    • Terminal count: 65,535
    • Maximum rate frequency:
      • 50 KHz with one enabled channel
      • 20 KHz with 2-4 enabled channels
      • 10 KHz with 5-7 enabled channels
    • Minimum rate frequency: 0.5 Hz
    • Maximum count frequency: 50 kHz

    Ethernet Interface

    • Type: 10/100Base-T
    • Connector: RJ-45
    • Protocol: TCP/IP
    • Server Type: DHCP or Fixed IP

    Removable Memory

    • Type: MLC, pSLC, or SLC flash memory (with USB SD card reader); USB thumb drive
    • Required Format: FAT32

    ChannelStretch™ Operation

    • USB
      • Maximum number of units: 16
      • Maximum channel count: 128 analog, 112 digital
      • Channel skew between any 2 units: 10 µS, typical
      • Maximum supported throughput: ≥480 kHz
    • Ethernet
      • Maximum number of units: 16
      • Maximum channel count: 128 analog, 112 digital
      • Channel skew between any 2 units: 100 µS, typical
      • Maximum supported throughput: >50 kHz

    Power Consumption: 14 Watts

    Datasheet

    Related Products

    DATAQ, USB DAQ, DI-4108-E, DI-4108-U, Data Acquisition, Data Logger
    USB/Ethernet Data Acquisition and Data Logger DI-4108-E

    DI-4108 High Speed USB and Ethernet Data Acquisition System with Windaq software. Features 8 differential analog input channels  12 to 16-bit measurement resolution

    Product detail
    DATAQ, DAQ, DI-788, Industrial Data Acquisition
    Industrial Data Acquisition DI-788

    DATAQ Instruments announces model DI-788, a new 32-channel data acquisition system designed specifically to accept DI-8B style modular signal conditioners and based on our Third...

    Product detail
    DATAQ, USB DAQ, DI-4718B-E, DI-4718B-U, DI-4718, Data Acquisition Data Logger
    USB Data Acquisition and Data Logger DI-4718B-U

    DI-4718B-U USB Data Logger and Data Acquisition System for applications that require signal conditioning. Install up to eight DI-8B isolated signal conditioning modules

    Product detail
    DATAQ, USB DAQ, DI-4208, DI-4208-U, DI-4208-E, Data Acquisition Data Logger
    USB/Ethernet Data Acquisition and Data Logger DI-4208-E

    DI-4208 High Speed USB and Ethernet Data Acquisition System with Windaq software. Features 8 differential analog input channels, 12 to 16-bit measurement resolution

    Product detail