Features:
- High precision current and voltage source ± 52 mA, ± 30 V
Option: ± 22 mA, ± 60 V - Precision simulation for all conventional thermocouple types (optional)
- Error limit 0.003 % Rdg.
- Standard with RS232 and IEEE488 interface, USB and Ethernet (optional)
- Current “SINK”
Specifications
Voltage source
Range ± | Resolution | Error limits at 23 °C ± of reading | TC with resp. to 23 °C |
30 V | 0.1 mV | 0.003 % (bis ± 4.5 V) + 200 µV (> ± 4.5 V) + 1.1 mV | 8 ppm/K + 10 µV/K |
3 V | 10 µV | 0.003 % (bis ± 450 mV) + 20 µV (> ± 450 mV) + 110 µV | 8 ppm/K + 1 µV/K |
300 mV | 1 µV | 0.003 % (bis ± 45 mV) + 3 µV (> ± 45 mV) + 11 µ | 8 ppm/K + 0.35 µV/K |
Option: 60 V
Range ± | Resolution | Error limits at 23 °C ± of reading | TC with resp. to 23 °C |
60 V | 0.2 mV | 0.003 % (bis ± 9 V) + 500 µV (> ± 9 V) + 2.2 mV | 8 ppm/K + 10 µV/K |
Current source
Range ± | Resolution | Error limits at 23 °C ± of reading | TC with resp. to 23 °C |
52 mA (22 mA | 200 nA | 0.007 % (bis ± 7.5 mA) + 0.6 µA (> ± 7.5 mA) + 3 µA | 10 ppm/K + 10 nA/K |
Option: Thermocouple simulation
Model | Range | Error (K) |
R | – 50.0 °C … 1768 °C | 0.4 (+ 250 … 1768 °C |
S | – 50.0 °C … 1768 °C | 0.4 (+ 350 … 1768 °C) |
B | 0.0 °C … 1820 °C | 0.5 (+ 800 … 1820 °C) |
J | – 210 °C … 1200 °C | 0.2 (- 170 … 400 °C) |
T | – 270 °C … 400 °C | 0.2 (- 220 … 1000 °C) |
E | – 270 °C … 1372 °C | 0.2 (- 220 … 1000 °C) |
K | – 270 °C … 1000 °C | 0.1 (- 50 … 800 °C) |
U | – 200 °C … 600 °C | 0.3 (- 100 … 600 °C) |
L | – 200 °C … 900 °C | 0.2 (- 100 … 750 °C) |
N | – 270 °C … 1300 °C | 0.2 (-120 … 1200 °C) |
M | – 50 °C … 1410 °C | 0.1 (- 50 … 900 °C) |
C | 0.0 °C … 2315 °C | 0.2 (+ 100 … 900 °C) |
D | 0.0 °C … 2315 °C | 0.2 (300 … 1100 °C) |
G2 | 0.0 °C … 2315 °C | 0.3 (300 … 2100 °C |
Q&A
Related Products
The resistance decade allows a direct choice of resistances between 1 Ω and 1.2 MΩ. Depending on the ohm value the resolution can be chosen...
MOBILE HIGH-PRECISION CALIBRATOR AND TESTING DEVICE FOR MECHANICAL AND ELECTRICAL MEASUREMENTS
DIGISTANT® 4420 allows the simulation and measurement of voltages, currents, temperatures and resistances.
HIGH-PRECISION MULTIFUNCTION CALIBRATOR FOR VOLTAGE, CURRENT, THERMOCOUPLES, RTDS, RESISTANCE AND FREQUENCY